thermosept® NDR

Combined disinfection and cleaning of non-invasive medical devices within chemothermal reprocessing

Our Plus

  • combines powerful cleaning with effective disinfection
  • aldehyde-free
  • pH-neutral and particularly gentle on materials
  • complies with the requirements of the working group "Decontamination systems for bedsteads and trolleys" (AK-BWA)
  • for use in bed and container washers

Downloads

Fields of Application

For chemothermal cleaning and disinfection of bedsteads, trolleys, containers, nightstands and operating tables in all common washers.






Instructions for use

Microbiological efficacy:
Standard concentration: With the parameters 1 % - 55 °C - 2 min., the product is bactericidal, yeasticidal and virucidal against enveloped viruses.
In order to avoid undesired interactions, use thermosept® NDR only together with thermosept® BSK. Both products are compatible. Thus, the ability to mix, as it inevitably occurs in many washers, is given. Using thermosept® BSK ensures spot-free drying of the washable items. The diluting effect occurring in the circulating pumps - due to the rinse aid solution - must be avoided by means of accordingly adjusted re-dosing. Daily exchange of the disinfecting and cleaning agent is required for hygienic reasons and to avoid loss of efficacy. After draining off the used disinfecting cleaning agent, the tank must be cleaned or rinsed.

Microbiological efficacy

EfficacyConcentrationContact time
bactericidal
EN13727, EN14561
- dirty conditions
1 % (10 ml/l)
55 °C
2 min.
yeasticidal
EN13624, EN14562
- dirty conditions
1 % (10 ml/l)
55 °C
2 min.
virucidal against enveloped viruses
EN14476, EN17111
- dirty conditions
1 % (10 ml/l)
55 °C
2 min.

Product data

Composition:
100 g of thermosept® NDR contains: 4.0 g dimethyldioctylammonium chloride, 3.0 g Phenoxypropanol





Chemical-physical data

Color:colourless
Viscosity, dynamic:No data available
Form:liquid
Flash point:> 100 °C / Method : ISO 2719
Density:ca. 1,00 g/cm3 / 20 °C
pH:ca. 6 / 20 °C

Special advice

Always read the label and product information before use.
For professional use only.
Any serious incidents related to the product are to be reported to the manufacturer and competent authority.
Please note recommendations of machine and medical device manufacturers.
Keep container tightly closed. Protect from frost, heat and sunlight. Store at room temperature in the original container.





Information for order

ItemDelivery formItem no.
thermosept NDR -D- 200 kg FA1/drum(s)on request
thermosept NDR -D- 20 l KA1/Canisteron request
These products are not available in every country. For more information please contact our local subsidiary or distributor.

Environmental information

schülke manufactures products economically and with advanced, safe and environmentally friendly production processes while at the same time maintaining out high quality standards.
CE 0297

Privacy Policy

We use analysis methods (e. g. cookies) to measure how often our site is visited and how it is used. In addition, we use cookies to link your page visits and website use with your customer data stored in our CRM in order to be able to address you individually, i.e. based on interests and usage. We embed third-party content from other providers (e. g. videos) on our site. We have no influence on further data processing and any tracking by the third party provider.
In this context, we also use service providers in third countries outside the EU that do not have an adequate level of data protection, which harbors the following risks: Access by authorities without informing the data subject, no data subject rights, no legal remedies, loss of control.
With your settings you consent to the processes described above. You can revoke your consent with effect for the future. You can find more information in our privacy policy.