Triple enzymatic cleaning of flexible endoscopes and surgical instruments of all types

Our Plus

  • turns milky at 32° C to keep temperature in optimal cleaning range (reversible process) and to avoid protein coagulation
  • safe in ultrasonic baths, for endoscopes, patients and users
  • faster action and more penetration
  • non-ionic surfactants form less foam
  • neutral pH makes it compatible to plastic, rubber and metals
  • protects from corrosions and damage

Downloads

Fields of Application

gigazyme® is an enzymatic cleaning agent designed specifically for the manual and semi-automatic cleaning of endoscopes and for the manual cleaning of surgical instruments.The ideal combination of high-quality enzymes (Protease = cutting proteins, Lipase = cutting fat, Amylase = cutting starch) and an innovative system of sufactants allows excellent cleaning of thermostable and thermolabile instruments of any kind. gigazyme® is also ideal for use in an ultrasonic bath.

Product Profile

gigazyme® is a triple-enzyme cleaner designed specifically for the manual and semi-automatic cleaning of endoscopes and for the manual cleaning of surgical instruments.

Instructions for use

gigazymeTM is a concentrate that can be diluted with cold water as per below mentioned instructions. For cleaning, immerse the instruments and endoscopes into stock solution in such a way that the surface is completely moistened and air escapes out of the lumen. After cleaning, rinse the instrument thoroughly.

In order to keep the temperature in optimal cleaning range, add cold water once solution starts turning milky. Renew the stock solution at least once a day depending on the degree of soiling.

gigazymeTM can be used in a range of 0.5% up to 10% concentration. To prepare 1 liter of 1% stock solution, add 10ml gigazymeTM to 990ml water.

Normal soiling:
gigazymeTM can be used in the concentration range of 0.5% to 1%.

Heavy soiling:
gigazymeTM can be used up to a concentration range of 10%

Product data

Composition:
5-15% non-ionic surfactants, enzymes (amylase, lipase & protease)

Special advice

Always read the label and product information sheet before use.

This product is intended for use by professionals only.

Any serious incidents associated with the product should be reported to the manufacturer and competent authority.

For optimal results, the temperature of the cleaning solution should be kept below 35°C. Failure to do so may cause blood to coagulate, particularly when instruments and endoscopes contaminated with blood are cleaned in an ultrasonic bath. The solution becomes turbid when it reaches approximately 32°C, indicating that the water temperature is too high. This turbidity does not have a negative impact on cleaning performance.

Do not mix gigazyme® with other products.

Information for order

ItemDelivery formItem no.
gigazyme 2 L6/Carton70003409


Environmental information



CE

Privacy Policy

We use analysis methods (e. g. cookies) to measure how often our site is visited and how it is used.
We embed third-party content from other providers (e. g. videos) on our site. We have no influence on further data processing and any tracking by the third party provider.
In this context, we also use service providers in third countries outside the EU that do not have an adequate level of data protection, which harbors the following risks: Access by authorities without informing the data subject, no data subject rights, no legal remedies, loss of control.
With your settings you consent to the processes described above. You can revoke your consent with effect for the future. You can find more information in our privacy policy.